Description
| Model: SONIMIX 2106 | 2106-10 | 2106-16 | 2106-64 | 2106-128 | 2106-256 | 2106-1024 |
| General data | ||||||
| Dilution points | 10 | 16 | 64 | 128 | 256 | 1024 |
| Dilution principle | Sonic nozzles (according ISO 6145-6) and high precision pressure regulators | |||||
|
Dilution gas line flow accuracy |
< 0.25 % of set point | |||||
| Source gas line flow accuracy | < 0.25 % | |||||
| Repeatability on flow | < 0.1 % | |||||
| Accuracy on concentration (rel) | < 0.5% , achieved on the highest dilution ratio | |||||
| Repeatability on concentration | Better than 0.14% relative | |||||
| Response time | < 15 sec | |||||
| Auto Diagnostic | Diagnostic of pressure regulator, valve opening, nozzle functioning, drift | |||||
| Factory calibration | By laminar flow with uncertainty < 0.2% of reading | |||||
| Warm up time | No warm up time | |||||
| Operating temperature | 15°C – 40°C | |||||
| Dimensions | 19″ 3HE/84 TE , 500mm deep
portable casing in option |
|||||
| Net weight | 15 – 20 Kg | |||||
| Gas inlet requirements | ||||||
| Dilution gas | Requires dry Air, Nitrogen or other gas, 2 ports | |||||
| Dew point | -40°C | |||||
| Source gas | compatible with corrosive gases up to 10% (if option installed)
up to 12 ports |
|||||
| Pressure | 3 ± 0.3 bar rel | |||||
| Fittings | 3 x Swagelok 1/4″ (Air, Nitrogen, Span gas)optional up to 3 x 4 Swagelok 1/4″ per instrument | |||||
| Gas outlet specifications | ||||||
| Flow | 2.5 ± 0.25 L/min , adjustable by by-pass valve in front of the instrument
other flow rate upon request |
|||||
| Pressure | up to 1 bar rel (other pressure upon request) | |||||
| Fittings | 4 x Swagelok 1/4″ (Mixture, Waste, Purge) | |||||
| Communication | ||||||
| RS 232 | Standard | |||||
| Ethernet TCP/IP | Standard | |||||
| USB | Standard | |||||
| WiFi | Optional | |||||
| Electrical data | ||||||
| Power supply voltage | 230V/50-60Hz
115V/60Hz 100V/50-60Hz |
|||||
| Installed power (Max) | 150W | |||||
| NOx converter tester (option) | ||||||
| NO2 range | 15 to 50 ppm / 30 to 100 ppm / 300 to 1000 ppm, all at 2.2 NL/min | |||||
| Stabilization time | Less than 1 min (once the device has warmed up) | |||||
| Inlet fittings | 2 x Swagelok (NO, air) 1/4″ | |||||
| Outlet fittings | 2 x Swagelok (outlet, purge) 1/4″ | |||||