AFS® D Water Purification System

AFS® D Water Purification System

0.00

CLRW quality

AFS® D Systems provide a cost-effective, low maintenance water purification solution for analyzers with degassed pure water needs up to 320 L daily. AFS® D systems combine complementary water purification techniques; Complete pretreatment, advanced reverse osmosis (RO), Q-Gard® polishing pack, exposure to a bactericidal UV lamp, degassing, and final 0.22 µm filtration ensure a reliable and consistent water quality. Water produced by the systems complies with the Clinical and Laboratory Standards Institute® (CLSI) guideline for clinical laboratory reagent water (CLRW). Depending on daily volume needs, different production flow rates are available (8 and 16 L/h).

The systems are designed to fulfill accreditation needs (such as CAP™ 15189 accreditation to the ISO® 15189:2007 standard), and have full monitoring and water quality archiving capabilities that allow storage of up to six months of data. Activation of Millitrack® Essential e-Solution allows easy, rapid access to this information and also enhances data management control and provides remote system access.

User friendly & easy maintenance

Robust AFS® D systems require little maintenance, which means less analyzer downtime. Systems have a small footprint and can be installed wherever it is convenient in the laboratory: on a wall, on or under the bench, or on a cart. AFS® D systems are customizable with a wide range of options and accessories, including reservoirs, a sanitization kit and an aeration kit. Optimized lifetimes for the systems’ pretreatment and polishing packs, as well as low feed water consumption make AFS® D systems a cost-effective choice.

Categories: Lab water purification system

Description

Water Specifications

AFS® D Essential CLRW water (CLSI Clinical Laboratory Reagent Water)

Resistivity at 25 °C > 15 MΩ•cm
TOC Typically < 30 ppb
Bacteria < 1 CFU/mL
Silica < 5 µg/L
Dissolved oxygen Typically 6 – 8 ppm
System Permeate Flow Rate Peak Dispense Rate (to atmosphere) Dispensing pressure to analyzer
AFS® 8D 8 L/h Up to 0.6 L/min* 1 bar ± 10 % (13-16 psi)*
AFS® 16D 16 L/h Up to 0.6 L/min* 1 bar ± 10 % (13-16 psi)*
* Dispensing to atmosphere (without any flow restriction due to the analyzer)

Feed water

Feed water quality Potable
Feed water pressure 1 – 6 bar
Feed water temperature 5 – 35 °C
Minimum flow rate feed > 5 L min at 2 bar
Total chlorine * < 3 ppm
Langelier Saturation Index < 0.3
Maximum Fouling Index (SDI) < 12
Tap water connection ½” Gaz M
* For feed water with total chlorine concentration between 1 ppm and 3 ppm, Merck Millipore suggests the use of an external PrePak.

System Specifications

Dimensions (H x W x D) 470 x 268 x 426 mm (18.5 x 10.6 x 16.8 in)
Operating weight < 22.2 kg (49 lb)
Electric power supply voltage 100 – 240 V +/- 10%
Electric power supply frequency 50 – 60 Hz +/- 10%
Power consumption 200 W or 250 VA
AFS® D main unit data connection Ethernet (RJ45)