Description
|
Models: Sonimix 4001 |
6800 17 000 standard range |
6800 17 001 extended range |
| General data | ||
|
Range |
From 10 to 1000 ppb in 1 ppb steps @ 3 L/min |
From 10ppb to 10 ppm @ 3 L/min + 10ppm to 35 ppm @ 0.9 L/min |
|
Dilution principle |
Sonic nozzles and high precision pressure regulators |
|
|
Dilution air |
Dry air from internal zero air (option) or external dry air source |
|
|
Generated gas flow |
3000 ml/min for range 1 |
900ml/min for range 2 |
|
Accuracy |
10 ppb to 100 ppb : +/- 2 ppb >100 ppb : 1.5% relative |
|
|
Ozone Repeatability |
Better than 1% relative |
|
|
Warm up time |
< 45 min |
|
|
Stabilization time |
< 15 min |
|
|
O3 regulation |
Measure of the UV light absorption of the generated ozone by UV reference photometer. (US EPA method and Beer Lambert law). Including atmospheric pressure and room temperature corrections |
|
|
Dimensions |
19cmx36cm, 50cm deep |
|
|
Net weight |
15 Kg |
|
|
UV Section |
||
|
UV lamp life time |
About 10’000 hours (with a mean concentration of 1 ppm) |
|
|
Zero Air section |
||
|
Life Time |
Annual preventive maintenance |
|
|
Quality |
O3, NO and NO2 < 1 ppb – H2O < -40°C dew point |
|
|
Communication |
||
|
Outlet Port |
Teflon 1/4‘’ Swagelok fitting |
|
|
Electrical data |
||
|
Power supply voltage |
230V/50-60Hz – 117V/60Hz – 100V/60Hz |
|